RESULTS: Five (8%) of the implants were revised because of aseptic loosening, 3% of the implants fractured, and 3% were revised because of infection. Six limbs (10%) required amputation because of local tumor recurrence. On radiographs, osteointegration into the collar was seen to have occurred in 70% of the patients and did not correlate with sex, age, diagnosis, or length of time postoperatively. Histological analysis showed mature lamellar bone within the grooves of the hydroxyapatite-coated collar, and bone was observed in direct contact with the hydroxyapatite coating. Extracortical bone failed to make direct contact with the surface of the implants manufactured without a collar.
CONCLUSIONS: The use of cemented distal femoral massive bone tumor prostheses with a hydroxyapatite-coated collar located at the shoulder of the implant was followed by a low (8%) rate of revision due to aseptic loosening. The use of hydroxyapatite grooved collars may lead to osteointegration of the implant shoulder (collar) and may reduce the rate of aseptic loosening.
LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.
PMID: 24005197 [PubMed - in process]