RESULTS: The mean amounts of proximal, two-dimensional, and three-dimensional head penetration between 1 week and 5 years were 0.018, 0.071, and 0.149 mm, respectively. The mean proximal, two-dimensional, and three-dimensional wear rates calculated between 1 year and 5 years were all less than 0.001 mm/year with no patient recording a wear rate of more than 0.040 mm/year.
CONCLUSIONS: The head penetration of a second-generation XLPE liner remained low at 5 years and the wear rate calculated after the first year was low in all directions. This low level of wear remains encouraging for the future clinical performance of this material.
LEVEL OF EVIDENCE: Level IV, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.
PMID: 23893361 [PubMed - as supplied by publisher]