RESULTS: Loop securities with the conventional method versus the WCS method were as follows: 10.74 ± 4.20 N versus 6.90 ± 3.90 N for Weston, 21.25 ± 14.74 N versus 8.73 ± 3.35 N for Nicky, 26.14 ± 15.57 N versus 7.95 ± 4.23 N for Roeder, 42.67 ± 22.96 N versus 8.67 ± 4.33 N for SMC, 52.99 ± 21.36 N versus 18.25 ± 10.58 N for San Diego, and 89.27 ± 27.96 N versus 12.48 ± 3.40 N for Dines (P < .05 for each knot). All knots failed at significantly lower loads when the suture loop was not evenly tensioned. Cadaveric testing (SMC) resulted in a loop security of 5.53 ± 6.06 N, which was similar to the WCS setting.
CONCLUSIONS: The locking mechanism of the sliding knots is maintained when the suture loop is evenly tensioned at both post and non-post strands. When tension is not applied to the non-post strand side, the knots slide more easily and fail at lower loads than previously reported.
CLINICAL RELEVANCE: When surgeons tie locking sliding knots in single-row rotator cuff repair, they should be aware that the knots could fail at much lower loads than previously reported.

PMID: 23906277 [PubMed - in process]

Read more... http://www.ncbi.nlm.nih.gov/pubmed/23906277?dopt=Abstract