RESULTS: We evaluated this automated system's capacity to reproducibly capture GH translation in intact and pathologic shoulder conditions. A pair of shoulders (left and right) were tested during forward elevation at baseline, with a winged scapula, and after creation of a full thickness supraspinatus tear.
DISCUSSION: The system detected differences in GH translations as small as 0.5 mm between different conditions. For each, three consecutive trials were performed and demonstrated high reproducibility and high precision.

PMID: 23883431 [PubMed - in process]

Read more... http://www.ncbi.nlm.nih.gov/pubmed/23883431?dopt=Abstract