RESULTS: Wear rates decreased with increasing polyethylene thickness. Mean ± SD wear rates for the 1.9-, 3.9-, 5.9-, and 7.9-mm groups were 5.0 ± 0.5, 3.2 ± 0.3, 2.5 ± 1.1, and 2.2 ± 1.3 mm(3)/million cycles, respectively (p < 0.016). Calculated penetration rates were 0.015, 0.012, 0.011, and 0.010 mm/million cycles, respectively (p < 0.016). There were no catastrophic failures in any group.
CONCLUSIONS: Thinner polyethylene components demonstrated higher wear rates, although even the highest wear rate observed in the thinnest polyethylene specimen was lower than that commonly reported for noncrosslinked polyethylene components. While encouraging, these findings should be validated in vivo before clinical recommendations can be made.
PMID: 23861047 [PubMed - as supplied by publisher]